OMNIScanner User Guide
2 - 4
1. To set the reference, connect OMNI
Scanner
and OMNI
Remote
using the test cable and
adapter supplied with the scanner (as shown).
2. Turn OMNI
Scanner
on.
3. Press (Setup).
4. Select Set Reference.
5. On finding OMNI
Remote
, the scanner reports:
Acquiring Reference Data. A gauge will indicate the progress.
6. Once the reference values has been recorded, OMNI
Scanner
displays
the REMOTE INFO screen.
7. Press ESC to return to OMNI
Scanner
’s SETUP screen.
Project Selection
A project name must be selected for test result management. Autotests
and Cables were previously assigned to a specific project through the use
of the OMNI
Scanner
Configuration utility.
(See Chapter 9 - Uploading and
Printing for further information about OMNIScanner Configuration.)
The Project feature is very useful for grouping
test results and eases uploading and sorting of
Autotests stored in the scanner.
1. Use the é or ê to highlight Project: and
press ENTER.
2. Select the appropriate Project from the drop-
down list and press ENTER.
3. Use the é or ê to highlight Autotest: and
press ENTER.
4. Select the appropriate Autotest from the drop-
down list and press ENTER.
5. Use the é or ê to highlight Cable: and press
ENTER.